Every incident, injury and recall on file, from the national registries that publish them. Search a maker, a model, a town or a reference number — or narrow by what happened. Each record is the registry's own text, reproduced without rewriting, with the reference that locates it at the source. Registries count different things, so a total is a count of FILINGS and never a measure of how dangerous something is. How the record was collected.
Or jump to a maker
6 records match for Intuitive Surgical, Inc.
20101 registryClear
Narrow it down
What happened
Country
Year
Not graded by the source2010-04-21FDA MAUDE1669451da Vinci Surgical System
da Vinci Surgical System — Injury report filed with FDA (MAUDE). A MAUDE report does not establish that the device caused the outcome. Reported narrative: UROLOGIST CALLED TO REPORT PROBLEMS WITH ROBOTIC EQUIPMENT. THE UROLOGIST HAD TO CONVERT THE ROBOTIC PROSTATECTOMY CASE TO OPEN WHEN THE SYSTEM STOPPED WORKING. THE NEXT WEEK, THE SURGEON HAD TWO ROBOTIC PROSTATECTOMIES. DURING THE FIRST CASE THERE WERE 10 ALARMS WITH THE ERROR #20013, THAT SAID PRESS FAULT OVERRIDE BUTTON. DESPITE THIS PROBLEM THE FIRST CASE WENT AS PLANNED WITHOUT ANY OTHER ISSUES. DURING THE SECOND CASE THE SAME PROBLEM WITH THE FAULT ALARM OCCURRED THREE TIMES; ON THE THIRD TIME THE FAULT DID NOT OVERRIDE. THE MANUFACTURER WAS CONTACTED USING THE DAVINCI HELP NUMBER. WE WERE INSTRUCTED TO UNDOCK THE ROBOT COMPLETELY AND POWER DOWN. I WAS INSTRUCTED TO CHECK SOME CONNECTIONS ON THE CONSOLE, WHICH WERE FINE. THEN WE TURNED ON THE ROBOT AND STARTED AGAIN. WE HAD ANOTHER FAULT THAT WE COULD NOT OVERRIDE, SO I CALLED THE COMPANY AGAIN, HAD THE SURGEON PRESS THE EMERGENCY STOP BUTTON AND THEN FAULT OVERRIDE. THIS WAS TO RELEASE THE INSTRUMENTS' GRASPS ON THE TISSUE. AT THAT TIME THERE WAS A BLEEDER THAT HADN'T FULLY GOTTEN UNDER CONTROL BECAUSE THE ROBOT WAS FROZEN. THERE WERE THREE MORE FAULTS BEFORE WE HAD ANOTHER IRRECOVERABLE FAULT. THE SURGEON DECIDED TO COMPLET
cause not established1669451Intuitive Surgical, Inc.Read the record ↗Not graded by the source2010-03-26FDA MAUDE2955842-2010-00122da Vinci Surgical System
da Vinci Surgical System — Injury report filed with FDA (MAUDE). A MAUDE report does not establish that the device caused the outcome. Reported narrative: IT WAS REPORTED THAT POST A DA VINCI SI THYROIDECTOMY PROCEDURE, THE PATIENT EXPERIENCED NUMBNESS DUE TO RADIAL NERVE DAMAGE.
cause not established2955842-2010-00122Intuitive Surgical, Inc.Read the record ↗Not graded by the source2010-03-18FDA MAUDE2955842-2010-00156da Vinci Surgical System
da Vinci Surgical System — Injury report filed with FDA (MAUDE). A MAUDE report does not establish that the device caused the outcome. Reported narrative: IT WAS REPORTED THAT POST A DA VINCI MYOMECTOMY PROCEDURE, THE PATIENT EXPERIENCED NUMBNESS IN HER HANDS AND FINGERS AND SHARP PAIN AND TINGLING IN HER LEGS. PER THE SURGEON THAT PERFORMED THE OPERATION, THE MYOMECTOMY PROCEDURE TOOK LONGER THAN ANTICIPATED, APPROXIMATELY 5 HOURS, WITH THE PATIENT IN THE TRENDELENBURG POSITION. PER THE OPERATIVE REPORT THE PATIENT TOLERATED THE ENTIRE PROCEDURE WELL WITH GOOD HEMOSTASIS AND MINIMAL BLOOD LOSS.
cause not established2955842-2010-00156Intuitive Surgical, Inc.Read the record ↗Not graded by the source2010-03-10FDA MAUDE2955842-2010-00140da Vinci Surgical System
da Vinci Surgical System — Injury report filed with FDA (MAUDE). A MAUDE report does not establish that the device caused the outcome. Reported narrative: IT WAS REPORTED THAT APPROXIMATELY 45 MINUTES INTO A DA VINCI PROSTATECTOMY PROCEDURE, WHILE THE SURGEON WAS PERFORMING A DISSECTION, THE MONOPOLAR FOOT PETAL WAS PRESSED, HOWEVER, THE BIPOLAR ENERGY BECAME ACTIVATED. THE BIPOLAR INSTRUMENT WAS HOLDING TISSUE AND CREATED A SUPERFICIAL LACERATION TYPE BURN TO THE PATIENT'S BOWEL. A GENERAL SURGEON REVIEWED THE INJURY AND DETERMINED THAT THE BURN WAS NOT THROUGH AND THROUGH AND DID NOT REQUIRE REPAIR, HOWEVER, THE SURGEON PLACED ONE SUTURE IN THE PATIENT'S BOWEL AS A PRECAUTION. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED WITH THE SAME EQUIPMENT AND NO FURTHER INSTANCES OF THE EVENT WAS REPORTED.
cause not established2955842-2010-00140Intuitive Surgical, Inc.Read the record ↗Not graded by the source2010-02-05FDA MAUDE2955842-2010-00084da Vinci Surgical System
da Vinci Surgical System — Injury report filed with FDA (MAUDE). A MAUDE report does not establish that the device caused the outcome. Reported narrative: IT WAS REPORTED THAT DURING A DA VINCI SI PROSTATECTOMY PROCEDURE, THE PHYSICIAN AT THE SURGEON SIDE CONSOLE INTERMITTENTLY FELT A SHOCK IN HIS FINGERTIPS FROM THE RIGHT MASTER TOOL MANIPULATOR (MTMR). THE SURGEON SWITCHED TO THE SITE'S SECOND SURGEON SIDE CONSOLE TO COMPLETE THE PLANNED PROCEDURE AND NO FURTHER OCCURRENCES OF SHOCK WERE REPORTED. NO HARM TO THE PHYSICIAN OR PATIENT WAS REPORTED.
cause not established2955842-2010-00084Intuitive Surgical, Inc.Read the record ↗Not graded by the source2010-01-27FDA MAUDE2955842-2010-00060da Vinci Surgical System
da Vinci Surgical System — Injury report filed with FDA (MAUDE). A MAUDE report does not establish that the device caused the outcome. Reported narrative: IT WAS REPORTED THAT APPROXIMATELY 2.5 HOURS INTO A DA VINCI SI HYSTERECTOMY PROCEDURE, THE SURGEON BEGAN TO INITIATE A TOOL CHANGE TO SWITCH INSTRUMENTS. AT THIS TIME THE NURSE ASSISTANT WHO WAS RETRACTING THE PATIENT'S UTERUS, DROPPED THE UTERUS CAUSING THE SURGEON TO LOSE SITE OF THE TENNACULUM FORCEPS INSTRUMENT IN USE. THE PATIENT SIDE MANIPULATOR (PSM) ARM THAT THE TENNACULUM FORCEPS INSTRUMENT WAS INSTALLED ON BECAME UNRESPONSIVE. THE ISI REPRESENTATIVE ONSITE HAD THE SURIGCAL STAFF EMERGENCY STOP THE SYSTEM AND PERFORM A FAULT OVER-RIDE, HOWEVER, THE PSM CONTINUED TO BE UNRESPONSIVE. THE PATIENT'S UTERUS WAS MOVED, AND THE SURGICAL STAFF FOUND THAT THE TENNACULUM FORCEPS INSTRUMENT WAS CLOSED ON THE PATIENT'S RECTUM AND THEY WERE UNABLE TO RELEASE THE INSTRUMENT. WITH THE ASSISTANCE OF AN ISI TECHNICAL SUPPORT ENGINEER, THE INSTRUMENT WAS RELEASED AND REMOVED FROM THE PSM. A GENERAL SURGEON WAS REQUESTED TO ATTEND THE PROCEDURE AND ALTHOUGH THERE WERE NO VISIBLE HOLES, TEARS, OR PINCHES IN THE PATIENT'S RECTUM, HE PLACED A FEW SUTURES AS A PRECAUTION. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED WITH OUT ANY FURTHER PATIENT HARM, INJURY OR ADVERSE OUTCOME REPORTED.
cause not established2955842-2010-00060Intuitive Surgical, Inc.Read the record ↗