Find a safety record
Every incident, injury and recall on file, from the national registries that publish them. Search a maker, a model, a town or a reference number — or narrow by what happened. Each record is the registry's own text, reproduced without rewriting, with the reference that locates it at the source. Registries count different things, so a total is a count of FILINGS and never a measure of how dangerous something is. How the record was collected.
Or jump to a maker
1 record match
20221 registryClear
Narrow it down
What happened
Country
Year
Recall2022-01-01Health CanadaRA-63783Mazor X Robotic Guidance System
Mazor X Robotic Guidance System — Type II recall (initiated 2021-12-16) for software v5.0.1 issues: potential unintended movement of the robotic arm into restricted work-volume areas (emergency stop remained functional), and possible undetected loss of communication between the console computer and the surgical arm causing unexpected arm positioning. One customer complaint of unintended arm movement; zero complaints on the communication-loss issue.
cause not establishedRA-63783Read the record ↗